Setting up Measurement Environment

Connecting the E5250A Input

Figure 4-7

Connection of Triax-Dual Triax Adapter

To

To

GNDU

SMU INPUT

TRIAX-DUAL TRIAX

Adapter

To

SMU

From

 

GND

Agilent 16493H Cable

 

Agilent E5250A

SMU INPUT

AUX

1

3

5

2

4

6

By this connection, input signal for SMU INPUT connectors is as follows:

SMU INPUT Port Number

Input Signal

 

 

1, 3, or 5

Force

 

 

2, 4, or 6

Sense

 

 

If you make non-Kelvin connection for the E5250A outputs, use the odd input port only. Because non-Kelvin connection uses only Force signal for measurement.

Agilent E5250A User’s Guide, Edition 9

4-19

Page 89
Image 89
Agilent Technologies Agilent E5250A manual Connection of Triax-Dual Triax Adapter, SMU Input Port Number Input Signal