Controlling the E5250A

Switch Control Basics

Channel Configuration

The E5250A allows the following two channel configuration modes. The way to specify the channel_list depends on the channel configuration mode. You can set the channel configuration by using the [:ROUT]:FUNC command described in Chapter 7.

Normal

Configuration Default mode at Power On or *RST command.

Each card is considered independently. For each channel in channel_list, you specify actual card number and channel number (input port number and output port number).

Auto

Configuration To use this mode, the same type of card must be installed in the E5250A slots from slot number 1 continuously (slots 1,2; or slots 1,2,3; or slots 1,2,3,4).

The installed cards are treated as one card for the output ports. For example, if the E5252As are installed in slots 1,2,3, you can specify output ports 1 to 36.

For each channel in channel_list, card number is always 0. For channel number, you specify the input port number and output port number depending on the type of card.

For more information, refer to “E5252A Channel List” on page 5-15 and “E5255A Channel List” on page 5-19.

5-8

Agilent E5250A User’s Guide, Edition 9