Executing Sample Programs

 

HCI Measurement Program

 

If you don't want program to pause, change the parameter in SP_MUX_A program to

NOTE

 

Pause_to_save=0 before you run the SP_MUX_A program. The program

 

automatically extracts Tdc, saves result data, and stops.

5.All calculated data is saved to ANAHCI file on computer's disk drive, which is an ASCII file. The data is also listed on the 4155/4156's screen. After analyzing, saving, and listing the data, the 4155/4156 is initialized.

For example, in the case that you selected Idlin softkey in step 3, the following is saved to the file:

Number of devices

Vdstr, Vgstr, Gate length, Gate width

Idlin

Device ,Validity ,Tdc_idlin

First device number, 0 or 1, Extracted Tdc for the device

:

:

:

Last device number, 0 or 1, Extracted Tdc for the device

Averaged Tdc_idlin

Calculated average Tdc

For the above, 0 means valid and 1 means invalid.

Following is an actual example for this case:

4

5, 1.95, 1.E-6, 1.E-5 Idlin

Device Validity Tdc_idlin 1, 0, 835.5786 2, 0, 3401.432 3, 0, 6269.047 4, 0, 24366.79 Averaged Tdc_idlin 8718.2119

9-34

Agilent E5250A User’s Guide, Edition 9

Page 290
Image 290