7-16 Agilent E5250A User’s Guide, Edition 9
Command Reference
:DIAGnostic:TEST:CARD:CLEar
:DIAGnostic:TEST:CARD:CLEar
This command clears the relay test result (pass/fail result) of the specified card.
Syntax :DIAGnostic:TEST:CARD:CLEar card_number
Example OUTPUT @Hp5250;":DIAG:TEST:CARD:CLE 1"
:DIAGnostic:TEST:CARD[:EXECute]?
This command executes the relay test for specified card, then returns the pass/fail
result. For more info, see Chapter 3.
NOTE You must attach the relay test adapter before executing this command.
This command changes all settings of card to same state as after executing
:SYST:CPON command.
Syntax :DIAGnostic:TEST:CARD[:EXECute]? card_number
Query response test_ result <newline><^END>
1 : Fail card exists
0 : Pass
Example OUTPUT @Hp5250;":DIAG:TEST:CARD? ALL"
Parameter Explanation
card_number card number: 1, 2, 3, 4, or ALL
Parameter Explanation
card_number card number: 1, 2, 3, 4, or ALL