Executing Sample Programs

Vth and Capacitance Measurement Program

To change the capacitance measurement setup

You can EDIT the program to change the following parameters in the "Constant Definition" block of the Meas_c subprogram.

Parameter

Description

Default

 

 

 

Test_comm$

Comment for the measurement.

"Capacitance

 

 

Measurement"

 

 

 

Capa_max

The maximum capacitance limit in F.

2.0E-11

 

 

 

Capa_min

The minimum capacitance limit in F.

8.0E-12

 

 

 

Sig_volt

Signal level for C measurement in V.

0.03

 

 

 

Meas_freq

Frequency for C measurement in Hz.

1.0E+5

 

 

 

Cable_len

Length (in meters) of 16048D/E cable.

1

 

 

 

Clen_hptriax

Length (in meters) of 16494A cable

1.5

 

connected from E5252A output to

 

 

connector plate.

 

 

 

 

Clen_usrtriax_h

Length (in meters) of triaxial cable

0

 

connected from connector plate to

 

 

DUT (High terminal of capacitor).

 

Clen_usrtriax_l

Length (in meters) of triaxial cable

0

 

connected from connector plate to

 

 

DUT (Low terminal of capacitor).

 

Clen_usrcoax_h

Length (in meters) of coaxial cable

1.0

 

connected from connector plate to

 

 

DUT (High terminal of capacitor).

 

Clen_usrcoax_l

Length (in meters) of coaxial cable

1.0

 

connected from connector plate to

 

 

DUT (Low terminal of capacitor).

 

Channel$

E5250A channel_list to use.

"(@10905,11006)"

 

 

 

To change the E5250A input port/output port connections, change the Channel$ parameter. And if you want to change the E5250A connection functions (Channel Configuration mode, etc.), modify the Set5250init subprogram.

9-14

Agilent E5250A User’s Guide, Edition 9

Page 270