Specifications

Agilent E5252A 10x12 Matrix Switch

Agilent E5252A 10x12 Matrix Switch

Switch Specifications

These are specifications when the E5252A is installed in mainframe. In the table, the conditions are as follows:

A: 23 °C ± 5 °C, 5 % to 60 % RH

B: 5 °C to 18 °C and 28 °C to 40 °C, 5 % to 60 % RH

C: 28 °C to 40 °C, 60 % to 80 % RH

 

 

A

B

C

 

 

 

 

 

Max. Current Rating (A)

 

1.0

1.0

1.0

 

 

 

 

 

Max. Voltage Rating (V)

 

 

 

 

 

 

 

Channel to Guard

 

200

200

200

 

 

 

 

 

Channel to Channel

 

300

300

300

 

 

 

 

 

Closed Channel Residual Resistance (Ω)

 

 

 

 

 

 

 

Low Leakage I-V Port

 

0.6

0.6

0.6

 

 

 

 

 

General I-V Port

 

1.0

1.0

1.0

 

 

 

 

 

C-V, HF Port

 

1.0

1.0

1.0

 

 

 

 

 

Channel Isolation Resistance (Ω)

 

 

 

 

 

 

 

Low Leakage I-V Port

 

1013

5 ⋅ 1012

2 ⋅ 1012

General I-V Port

 

1012

5 ⋅ 1011

2 ⋅ 1011

C-V, HF Port

 

109

5 ⋅ 108

2 ⋅ 108

10-4