Executing Self-Test and Leak Test

 

 

 

 

 

Using the Self-Test Utility

 

 

To Execute Leak Test

 

 

 

 

Leak Test is for checking leakage current of the E5250A with plug-in cards

NOTE

 

 

 

installed. If leak test result is too large for your application, contact nearest Agilent

 

 

Technologies Service Center. Leak test does not verify specifications of the E5250A

 

 

and installed cards.

 

 

 

 

 

The Leak Test applies 10 Vdc to the specified input port of the E5250A, and

 

 

 

 

measures the leakage current. The test limits are shown in Table 3-4.

 

 

 

You specify an input port of the E5250A, and specify the 4155/4156 SMU to use.

 

 

The Leak Test automatically connects the specified input port to each output port

 

 

consecutively, and uses the 4155/4156 SMU to measure the leakage current for each

 

 

channel.

 

 

 

 

 

 

You need to execute this test for each input port.

 

Table 3-4

 

Test Limits of Leak Test

 

 

 

 

 

 

 

 

 

 

 

 

Card

Instrument

Connection

Input Port

Test Limit

 

 

 

 

 

 

 

 

 

E5252A

4155

non-Kelvin

IV1, IV2

±4.12 pA

 

 

 

 

 

 

 

 

 

 

 

 

IV3, IV4, IV5, IV6

±1.01 nA

 

 

 

 

 

 

 

 

 

 

4156

non-Kelvin

IV1, IV2

±1.14 pA

 

 

 

 

 

 

 

 

 

 

 

 

IV3, IV4, IV5, IV6

±1.01 nA

 

 

 

 

 

 

 

 

 

 

 

Kelvin

IV1, IV2

±2.24 pA

 

 

 

 

 

 

 

 

 

 

 

 

IV3, IV4, IV5, IV6

±2.02 nA

 

 

 

 

 

 

 

 

 

E5255A

4155

non-Kelvin

IV1 to IV6

±4.12 pA

 

 

 

 

 

 

 

 

 

 

4156

non-Kelvin

IV1 to IV6

±1.14 pA

 

 

 

 

 

 

 

Agilent E5250A User’s Guide, Edition 9