Agilent E5250A User’s Guide, Edition 9 7-45
Command Reference
:SYSTem:CPON
:SYSTem:CPON
This command resets the specified card to the power-on state. If the parameter is
ALL, all addressable cards are reset to the power-on state. Status after this command
is shown below:
Channel Configuration: No Change
Connection Rule: Free
Connection Sequence: Break Before M ake
Bias Mode: Off
Bias Port: E5252A : 10
E5255A : None
Bias Enabled Channels: All channels (output ports) are bias enabled
Couple Port Mode: Off
Couple Port: No Couple Ports
Card Channel Status: All relays open in present channel configuration mode
Self-Test Result: No Change
For the device reset state, see *RST.
Syntax :SYSTem:CPON card_number
Example OUTPUT @Hp5250;":SYST:CPON ALL"
Parameter Explanation
card_number card number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL