Setting up Measurement Environment

Connecting the E5250A Input

To Connect Bias Source to the E5255A Bias Input

The E5255A BIAS INPUT connector is a BNC-type connector.

Use BNC cable to connect the bias source to the BIAS INPUT of the E5255A.

If output terminal of the bias source is a banana plug, use a Dual Banana Plug to

BNC Adapter (Agilent part number 1251-2277).

To Connect GNDU to the E5250A Input

Ground Unit (GNDU) of Agilent 4142B Modular DC Source/Monitor or Agilent 41501 Expander (for Agilent 4155/4156) is usually connected to GNDU connector on the connector plate by using Agilent 16493H GNDU cable. This connection allows current over 1A.

If you want to connect the GNDU through the E5250A to connector plate, test fixture or DUT, connect the GNDU to the E5250A input as shown below.

CAUTION

Make sure current to GNDU is less than 1 A if you connect GNDU to the E5250A.

 

The maximum input current of the E5250A is 1A.

 

Required Parts:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Agilent

Quantity

Description

 

 

 

Model/Part No.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16493H

1

GNDU Cable

 

 

 

 

 

 

 

 

 

E5250-60044

1

Triaxial to Dual Triaxial Adapter

 

 

 

 

 

 

 

Procedure:

1.Connect the adapter to the E5250A SMU INPUT connector (1&2, 3&4 or 5&6) as shown in Figure 4-7.

2.Connect the GNDU cable between the instrument GNDU connector and the triaxial connector on the adapter.

4-18

Agilent E5250A User’s Guide, Edition 9

Page 88
Image 88