WARNING

CAUTION

NOTE

Installation

Installing the E5250A

Installing the E5250A

This section describes how to install the E5250A.

“To Install Plug-in Card”

“To Install Blank Panel”

“To Configure E5255A”

“E5255A Configuration Examples”

“To Set GPIB Address”

“To Connect GPIB Cable”

To prevent electrical shock, turn off the mainframe and remove the power cable before starting the instruction.

Be careful about the module pins used for internal connection to the mainframe. The pins can be damaged easily.

Use clean handling and anti-static procedures when removing, configuring, and installing the modules. The modules contain components that can be damaged by static electricity.

Plug-in module is shipped from the factory after the specifications are confirmed. Agilent can guarantee that the modules will function and that the module performance is designed to meet its specifications. However, if you have any trouble, contact your nearest Agilent Technologies service center.

If you require that the modules be guaranteed to their specifications (for ISO compliance, etc.), then you must perform a calibration on the instrument (modules and mainframe together). For calibration, contact your nearest Agilent Technologies service center.

Agilent E5250A User’s Guide, Edition 9

2-7

Page 39
Image 39
Agilent Technologies Agilent E5250A manual Installing the E5250A