Executing Self-Test and Leak Test

Using the Self-Test Utility

To execute the Leak Test, do the following:

1.On MAIN MENU of the E5250A Self-Test Utility, select LEAK TEST softkey. The following message is displayed:

Input Agilent 4155/4156 GPIB address (default 817):?

2.To use default GPIB address 817 for the 4155/4156, just press Enter key. If not, type GPIB address of the 4155/4156, then press Enter key.

The LEAK TEST MENU with the following softkeys are displayed:

Softkey

Description

 

 

START TEST

For defining the Leak Test connection. Need to define the

 

E5250A input port and 4155/4156 SMU port.

 

 

MAIN MENU

Returns to MAIN MENU of the Self-Test Utility.

 

 

3. Select START TEST softkey. Following message/softkeys are displayed:

Select IV Port of Leak Test by softkey.

Softkey

Description

 

 

IVn (n=1 to 6)

For specifying the E5250A SMU INPUTn port for which

 

you want to perform the leak measurement. If only the

 

E5255As are installed, only IVn softkeys displayed are for

 

SMU INPUTn internally connected to the E5255A IV input.

 

 

CANCEL

Cancels the Leak Test execution.

 

 

3-16

Agilent E5250A User’s Guide, Edition 9