Setting up Measurement Environment

Mounting Connectors Directly

NOTE

To Check Interlock Circuit

 

If you use the 4155/4156, you can easily check the interlock circuit as follows:

 

1.

Connect the Intlk connector of the 4155/4156 to your interlock circuit.

 

2.

Press System front-panel key, then select CALIB/DIAG primary softkey to

 

 

display the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS page.

 

3.

In the CALIB/DIAG field, select DIAG secondary softkey.

 

4.

In the CATEGORY field, select I/O PERIPH secondary softkey.

 

5.

Move pointer to the 403 (INT.) Interlock LED field.

 

6.

Select EXECUTE secondary softkey.

 

7.

Confirm the following:

 

 

• LED turns on within 1 sec from when interlock circuit is shorted.

 

 

• LED turns off within 1 sec from when interlock circuit is open.

To stop the interlock test, select STOP secondary softkey.

4-12

Agilent E5250A User’s Guide, Edition 9

Page 82
Image 82
Agilent Technologies Agilent E5250A manual To Check Interlock Circuit