Programming the E5250A

Creating a Control Program

Table 6-1

Programming Summary

 

 

 

 

 

 

Description

Command

Default Setting

 

 

 

 

1

Defines Channel Configuration Mode

:ROUT:FUNC channel_config

NCON

 

 

 

 

2

Defines Connection Rule

:ROUT:CONN:RULE card_no, rule

ALL,FREE

 

 

 

 

3

Defines Connection Sequence

:ROUT:CONN:SEQ card_no,sequence

ALL,BBM

 

 

 

 

 

4

Using Bias Mode

 

 

 

 

 

 

 

 

Selects input Bias Port (for E5252A)

:ROUT:BIAS:PORT card_no, input_port_no

ALL,10

 

 

 

 

 

Bias enables all output ports

:ROUT:BIAS:CHAN:ENAB:CARD card_no

ALL

 

 

 

 

 

Bias enables specified output ports

:ROUT:BIAS:CHAN:ENAB:LIST channel_list

All enabled

 

 

 

 

 

Bias disables all output ports

:ROUT:BIAS:CHAN:DISA:CARD card_no

All enabled

 

 

 

 

 

Bias disables specified output ports

:ROUT:BIAS:CHAN:DISA:LIST channel_list

All enabled

 

 

 

 

 

Sets Bias Mode ON/OFF

:ROUT:BIAS:STAT card_no, state

ALL,OFF

 

 

 

 

5

Using Couple Port (for E5252A)

 

 

 

 

 

 

 

 

Sets up Couple Port

 

:ROUT:COUP:PORT card_no, 'input_port_no'

No couple ports

 

 

 

 

 

Sets Couple Port Mode ON/OFF

:ROUT:COUP:STAT card_no, state

ALL,OFF

 

 

 

 

6

Controls Relays on Cards

:ROUT:CLOS channel_list

All relays open

 

 

 

 

 

 

 

 

:ROUT:OPEN channel_list

All relays open

 

 

 

 

 

 

 

 

:ROUT:OPEN:CARD card_no

ALL

 

 

 

 

 

 

 

Parameter

Description

 

 

 

 

 

 

channel_config

NCON or ACON.

 

 

 

 

 

 

rule

FREE or SROU.

 

 

 

 

 

 

sequence

NSEQ, BBM or MBBR.

 

 

 

 

 

 

state

ON (1) or OFF (0).

 

 

 

 

 

 

channel_list

5 digit decimal number. See Chapter 5.

 

 

 

 

 

 

card_no

ALL or 0 or 1 to 4 are available.

 

 

 

 

 

 

input_port_no

1 to 10 are available for :ROUT:BIAS:PORT.

 

 

 

1, 3, 5, 7 or 9 are available for :ROUT:COUP:PORT.

 

 

For details about these parameters, refer to Chapter 5 and Chapter 7.

 

 

Bias Port and Couple Port cannot be set for the same input port number.

CAUTION

 

Agilent E5250A User’s Guide, Edition 9

6-7