Executing Self-Test and Leak Test

Using the Self-Test Utility

4.Select desired IVn softkey (n=1 to 6) for input port you want to test.

If the E5252A is installed in your E5250A, and the 4156 is used for the Relay Test, then when you press IV1, IV3, or IV5 softkey, the following message and softkeys are displayed:

Is IVn Port a Kelvin connection?

Where n=1, 3, or 5.

Softkey

Description

 

 

YES

Declares that the port specified by the IVn softkey is used

 

for the Kelvin connection.

 

 

NO

Declares that the port specified by the IVn softkey is used

 

for the non-Kelvin connection.

 

 

CANCEL

Cancels the Leak Test execution.

 

 

Select the YES or NO softkey.

The following message and the softkeys are displayed:

Select SMU channel for IVn leak measurement by softkey.

Where n=1 to 6.

Softkey

Description

 

 

SMUm

Declares that you will connect the 4155/4156 SMUm to the

(m= 1 to 6)

E5250A port specified by the IVn softkey. SMU5 and

 

SMU6 are displayed only if 5th and 6th SMU are installed

 

in the 41501 expander of the 4155/4156.

 

 

CANCEL

Cancels the Leak Test execution.

 

 

Agilent E5250A User’s Guide, Edition 9

3-17

Page 67
Image 67
Agilent Technologies Agilent E5250A manual Yes