6-20 Agilent E5250A User’s Guide, Edition 9
Programming the E5250A
Programming Examples
Program List: 10 ! Example: Using Couple Port
20 !
30 INTEGER Complete
40 DIM Channel$[100]
50 ! (Substrate,Gate,Drain,Source)
60 Channel$="(@101,202,303,505)"
70 !
80 ! Put Instrument Initialize Routine here
90 !
100 ASSIGN @Hp5250 TO 722
110 OUTPUT @Hp5250;"*RST"
120 OUTPUT @Hp5250;":ROUT:FUNC ACON"
130 OUTPUT @Hp5250;":ROUT:CONN:RULE ALL,SROU"
140 OUTPUT @Hp5250;":ROUT:CONN:SEQ ALL,BBM"
Line
Number Description
40 Declares data size of Channel$, which is used as channel_list
parameter.
60 Defines th e desired channel_list.
80 Here, insert the initialization routine for your measurement
instrument.
110 Resets the E5250A.
120 Sets the E5250A to Auto channel configuration mode.
130 Sets the connection rule to Single.
140 Sets the connection sequence to Break-Before-Make.