Programming the E5250A

Programming Examples

Program List:

10

! Example: Using Couple Port

 

20

!

 

 

30

INTEGER Complete

 

40

DIM Channel$[100]

 

50

! (Substrate,Gate,Drain,Source)

 

60

Channel$="(@101,202,303,505)"

 

70

!

 

 

80

! Put Instrument Initialize Routine here

 

90

!

 

 

100

ASSIGN @Hp5250 TO 722

 

110

OUTPUT @Hp5250;"*RST"

 

120

OUTPUT @Hp5250;":ROUT:FUNC ACON"

 

130

OUTPUT @Hp5250;":ROUT:CONN:RULE ALL,SROU"

 

140

OUTPUT @Hp5250;":ROUT:CONN:SEQ ALL,BBM"

 

 

 

 

 

 

Line

Description

 

Number

 

 

 

 

 

 

 

 

40

Declares data size of Channel$, which is used as channel_list

 

 

 

parameter.

 

 

 

 

 

 

60

Defines the desired channel_list.

 

 

 

 

 

 

80

Here, insert the initialization routine for your measurement

 

 

 

instrument.

 

 

 

 

 

 

110

Resets the E5250A.

 

 

 

 

 

 

120

Sets the E5250A to Auto channel configuration mode.

 

 

 

 

 

 

130

Sets the connection rule to Single.

 

 

 

 

 

 

140

Sets the connection sequence to Break-Before-Make.

 

 

 

 

6-20

Agilent E5250A User’s Guide, Edition 9

Page 156
Image 156