Agilent E5250A User’s Guide, Edition 9 8-11
VXIplug&play Driver
Driver Functions
hpe5250a_closeList
This function will connect the input ports to the output ports specified by the
channel list.
The parameter ’closechan_list’ is an array of integers with each integer representing
one channel. The last number of the ’closechan_list’ should be "0" (numeric zero)
to identify the end of the list. The maximum number of channels that can be
specified by the list is 100.
Syntax ViStatus _VI_FUNC hpe5250a_closeList(ViSession vi,
ViInt32_VI_FARclosechan_list[ ] );
Parameters vi Instrument handle returned from hpe5250a_init( ).
closechan_list[ ] Channel numbers to connect. 5 digits integer. ABCDE. where
A: card number, BC: input port number, DE: output port
number. Topzero(s) can be ignored. For example, if A=0,
BC=01, and DE=01, channel number should be 101 instead of
00101.