Error Messages

E5250A Specific Error Messages

E5250A Channel Related Errors

Error

Error Message and Description

Number

 

 

 

2000

Invalid card number

 

Wrong card number is specified in card number or channel list

 

parameter. Make sure card is properly installed in the correct

 

E5250A slot.

 

 

2001

Invalid channel number

 

Wrong channel number is specified in the channel list. Confirm the

 

channel number, the card configuration, and the configuration

 

mode of the E5250A.

 

 

2006

Command not supported on this card

 

Command was used that is not supported by the card.

 

 

2009

Too many channels in channel list

 

Too many channels are specified in the channel list. Maximum

 

number of channels:

 

• E5252A: number of cards ⋅ 120

 

• E5255A: number of cards ⋅ 48

 

 

2011

Empty channel list

 

No channel list is specified.

 

 

2012

Invalid channel range

 

Wrong channel list is defined. Confirm the syntax of the channel

 

list, the card configuration, and the configuration mode of the

 

E5250A.

 

 

Agilent E5250A User’s Guide, Edition 9

11-11

Page 315
Image 315