11-8 Agilent E5250A User’s Guide, Edition 9
Error Messages
Standard SCPI Error Messages
Device-Dependent Errors
-3XX errors indicate that an E5250A operation did not properly complete, possibly
due to an abnormal hardware or firmware condition. These negative codes are SCPI
defined. For the device-dependent positive error codes, see “E5250A Specific Error
Messages” on page 11-10. The positive codes are not SCPI defined.
For these errors, an error number and message are placed in the error queue, and bit3
of the Standard Event Status Register is set.
Error
Number Error Message and Description
-300 Device-specific error
Generic device-dependent error for the E5250A that cannot be
determined more specifically.
-311 Memory error
An error was detected in the E5250A's memory.
-350 Queue overflow
If error queue is full, this code is entered into the error queue instead
of the code that caused the error. This code indicates that there was
no room in the error queue, so an error occurred but was not
recorded.