Installation

Installing the E5250A

To Configure E5255A

The Agilent E5255A has three 2-input 8-output multiplexers. You can configure several multiplexers by modifying some internal connections. Component locations on the E5255A are shown in Figure 2-2, and the relationship of the multiplexer blocks and the components is shown in Table 2-1. Table 2-2 shows the initial setting of the E5255A internal connections. The setting does not depend on the slot in which the module is installed.

Table 2-1

Multiplexer Blocks and Components

 

 

 

 

 

 

 

 

 

 

 

 

Multiplexer

Resistor

BIASin

BIASout

IVin

IVout

DIP SW

 

block

holder

cable

connector

 

 

 

 

 

 

 

 

 

 

 

 

 

Block 1

holder 1

-

BIASout1

IVin1

IVout1

DIP SW1

 

Block 2

holder 2

BIASin2

BIASout2

IVin2

IVout2

DIP SW2

 

 

 

 

 

 

 

 

 

Block 3

holder 3

BIASin3

-

IVin3

IVout3

DIP SW3

Table 2-2

Agilent E5255A Initial Setup

 

 

 

 

 

 

 

 

 

 

Block number

IVin cable connection

BIASin/BIASout

Protective resistor

 

 

 

 

 

 

 

 

1

SMU1

BIAS INPUT 51

0 Ω

 

 

 

 

 

 

 

 

2

SMU2

BIAS INPUT 52

0 Ω

 

 

 

 

 

 

 

 

3

SMU3

BIAS INPUT 53

0 Ω

 

 

 

 

 

 

To Modify IV Input

IV input connection is defined by the connection of IVin cable and the setting of DIP switch shown in Figure 2-2. For actual examples of modifications, see “E5255A Configuration Examples”.

Each DIP switch has four bit switches. Bits 1 to 3 are used to specify which SMU input connector is connected to the IV input. Bit 4 of DIP switch is used to specify whether the protective resistors are 0 Ω or not. The DIP switches must be set properly after you complete the internal connection. Make sure you set the DIP switches correctly. If not, dangerous voltages may be applied to unexpected output connectors.