Controlling the E5250A

Using Virtual Front Panel Utility

The following softkeys are displayed on right side of the 4155/4156 screen.

Softkey

Description

 

 

SHOW

For confirming which cards are installed in each E5250A slot.

CONFIG

 

 

 

SETUP

For changing the VFP setup mode (Channel Configuration mode,

MODE

Bias Port, and Bias Mode state). When the setup mode is changed,

 

all setup data is cleared (that is, connection data is set so that all

 

relays are open).

 

 

DEFINE

For defining labels for the E5250A input ports to be used in

LABEL

MEMORY SETUP PANEL.

 

 

SETUP

For creating connection channels in the E5250A cards. When you

MEMORY

modify any setting of the setup memory, the E5250A connection

 

channels are changed in real time.

 

 

SAVE/

For saving/loading the setup data to/from diskette.

LOAD

 

 

 

QUIT

Quits the VFP utility.

 

 

Confirming the E5250A Card Configurations

Select SHOW CONFIG softkey. SHOW CONFIGURATION MENU is displayed on the 4155/4156 to show which card is installed in each E5250A slot.

If card is the E5252A, the following comment is displayed:

E5252A 10x12 Matrix Switch

If card is the E5255A, a comment similar to following example is displayed:

E5255A 24 (8x3) Channel Multiplexer

Block 1 : Input 1, No Resistance

Block 2 : Input 2, No Resistance

Block 3 : Input 3, No Resistance

For Input N, the N can be 1 to 6, and indicates which E5250A SMU input port is connected to the IV input of each multiplexer block. The next parameter indicates if resistor is used between input Bias Port and IV Input ports as follows:

No Resistance: no resistor used.

With Resistance: resistor is used.

5-28

Agilent E5250A User’s Guide, Edition 9

Page 122
Image 122
Agilent Technologies Agilent E5250A manual Confirming the E5250A Card Configurations, Mode