Programming the E5250A

Programming Examples

160 !

170 ! Put Bias Source Control Routine here.

180 !

190 OUTPUT @Hp5250;":ROUT:CLOS "&Channel$

200 OUTPUT @Hp5250;"*OPC?"

210 ENTER @Hp5250;Complete

220 !

230 ! Put Measurement Routine here

240 !

250 OUTPUT @Hp5250;":ROUT:BIAS:STAT ALL,OFF"

260 OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"

270 OUTPUT @Hp5250;"*OPC?"

280 ENTER @Hp5250;Complete

290 !

300 OUTPUT @Hp5250;"*RST"

310!

320END

Line

Description

Number

 

 

 

170

Here, insert your bias source control routine.

 

 

190

Disconnects all output ports of block 1 from the input Bias Port,

 

then connects the output ports to the IV input port.

 

 

200 to 210

Waits until the channel connections are completed.

 

 

230

Here, insert measurement routine for your measurement

 

instrument.

 

 

250

Turns off the Bias Mode.

 

 

260 to 280

Opens all relays, then connects all output ports of block 1 to the

 

input Bias Port again, and waits until the connections are

 

completed.

 

 

300

Resets the E5250A.

 

 

6-18

Agilent E5250A User’s Guide, Edition 9

Page 154
Image 154
Agilent Technologies manual Agilent E5250A User’s Guide, Edition