Executing Self-Test and Leak Test

Using the Self-Test Utility

To Start the Self-Test Utility

1.Before turning on the 4155/4156 and E5250A, connect the GPIB cable between the E5250A and the 4155/4156.

2.Turn on the 4155/4156 and E5250A.

3.Set the 4155/4156 to "SYSTEM CONTROLLER" mode on SYSTEM: MISCELLANEOUS page.

4.Insert the E5250A Program Disk into the 4155/4156 disk drive.

5.Press the Display key twice to display the "All IBASIC" screen. This key is located in the IBASIC area on the 4155/4156 front panel.

6.Select GET softkey on the 4155/4156 screen. The following is displayed:

GET "_"

7.Type SELFTEST between the double quotations of GET statement. Then press Enter key. Wait until the 4155/4156 finishes getting the file.

8.Press Run key, which is in the IBASIC area on the 4155/4156 front panel. The following message is displayed:

Input Agilent E5250A GPIB Address (default 722):?

9.To use default GPIB address 722, just press the Enter key. If not, type the GPIB address of your E5250A, then press Enter key. MAIN MENU of the E5250A Self-Test Utility and following softkeys are displayed on the 4155/4156:

Softkey

Description

 

 

SELF TEST

Displays the SELF TEST MENU.

 

 

LEAK TEST

Displays the LEAK TEST MENU.

 

 

QUIT

Quits the Self-Test Utility.

 

 

Agilent E5250A User’s Guide, Edition 9

3-11

Page 61