Executing Self-Test and Leak Test

Using the Self-Test Utility

Using the Self-Test Utility

The Self-Test Utility is a Self-Test program for the E5250A, and runs on the Agilent 4155/4156's built-in IBASIC controller.

Self-Test Utility can execute all Self-Test items listed in Table 3-1 and can also execute the Leak Test, which checks leakage current of the E5250A with installed cards. You execute the Self-Test items and Leak Test interactively (using softkeys) from the 4155/4156.

This section explains the operation of the Self-Test Utility.

“Requirements”

“To Start the Self-Test Utility”

“To Execute Self-Test”

“To Execute Leak Test”

Requirements

 

The following equipment is required to execute the Self-Test Utility:

Table 3-3

Required Equipment

 

 

 

 

 

 

 

 

Agilent

Description

Qty.

Usage

 

Model/Part No.

 

 

 

 

 

 

 

 

 

 

4155 or 4156

Semiconductor

1

Executing

 

 

Parameter Analyzer

 

Self-Test

 

 

 

 

Utility

 

10833A/B/C

GPIB Cable

1

 

 

 

 

 

 

 

 

E5250-17003

E5250A Program

1

 

 

 

Disk (furnished)

 

 

 

 

 

 

 

 

E5250A-301

Relay Test Adapter

1

Relay Test

 

(E5250-60002)

 

 

 

 

 

 

 

 

 

1250-0929

BNC Short Cap

3 for each E5255A

 

 

 

 

(furnished with the

 

 

 

 

E5255A)

 

 

 

 

 

 

3-10

Agilent E5250A User’s Guide, Edition 9

Page 60
Image 60