3-10 Agilent E5250A User’s Guide, Edition 9
Executing Self-Test and Leak Test
Using the Self-Test Utility
Using the Self-Test Utility
The Self-Test Utility is a Self-Test program for the E5250A, and runs on the Agilent
4155/4156's built-in IBASIC controller.
Self-Test Utility can execute all Self-Test items listed in Table 3-1 and can also
execute the Leak Test, which checks leakage current of the E5250A with installed
cards. You execute the Self-Test items and Leak Test interactively (using softkeys)
from the 4155/4156.
This section explains the operation of the Self-Test Utility.
• “Requirements”
“To Start the Self-Test Utility”
“To Execute Self-Test”
“To Execute Leak Test”

Requirements

The following equipment is required to execute the Self-Test Utility:
Table 3-3 Required Equipment
Agilent
Model/Part No. Description Qty. Usage
4155 or 4156 Semiconductor
Parameter Analyzer 1 Executing
Self-Test
Utility
10833A/B/C GPIB Cable 1
E5250-17003 E5250A Program
Disk (furnished) 1
E5250A-301
(E5250-60002) Relay Test Adapter 1 Relay Test
1250-0929 BNC Short Cap 3 for each E5255A
(furnished with the
E5255A)