Installation Installing the E5250A

Figure 2-2

Agilent E5255A Component Locations

Module pins

 

 

 

SMU input connectors

 

 

(1) (2) (4) (R)

 

 

 

 

BIT1 BIT2 BIT3 BIT4

0: OFF

SMU1

SMU2 SMU3 SMU4 SMU5 SMU6

 

1: ON

DIP SW1

IVin cable

DIP SW setting

 

 

connection

BIT1 BIT2

BIT3

IVout1

IVout2

SMU1

1

0

0

connector

connector

SMU2

0

1

0

IVin1

IVin2

SMU3

1

1

0

cable

cable

SMU4

0

0

1

 

 

SMU5

1

0

1

 

 

DIP SW2

IVout3 connector

DIP SW3

IVin3 cable

SMU6

0

1

1

Block 1

Block 2

 

 

Resistors

BIT4

 

 

 

 

0 ohm

0

 

 

Resistor holder 1

Resistor holder 2

other

1

 

 

 

 

Screws

 

 

 

 

BIASin2

(Total 48)

 

 

 

 

 

 

terminal

 

 

 

 

 

Resistors

 

 

 

BIASout1

(Total 24)

 

 

 

 

 

 

 

 

terminal

 

 

 

 

To BIAS INPUT 51

To BIAS INPUT 52

Block 3

Resistor holder 3

BIASin3 terminal

BIASout2 terminal

To BIAS INPUT 53

To Modify BIAS Input

Each multiplexer block has an independent BIAS input. You can change the BIAS input internally using the BIASin/BIASout terminals shown in Figure 2-2, or externally using the BIAS INPUT connectors. For actual examples of modifications, see “E5255A Configuration Examples”.

To Use Multiplexer Blocks on another E5255A

When a module is installed in a mainframe slot, six module pins are connected to the mainframe internal bus. So, if multiplexers on different E5255As use the same module pin, the IV inputs are connected together via the internal bus.

Agilent E5250A User’s Guide, Edition 9

2-11