7-40 Agilent E5250A User’s Guide, Edition 9
Command Reference
[:ROUTe]:OPEN:CARD
[:ROUTe]:OPEN:CARD
This command disconnects all input ports from all output ports for the specified
card. Then, if Bias Mode is ON, connects the input Bias Port to all bias enabled
output ports.
For details about the Bias Mode, refer to ":ROUTe:BIAS:STATe" in this chapter and
refer to Chapter 5.
Syntax [:ROUTe]:OPEN:CARD card_number
Example OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"
[:ROUTe]:OPEN[:LIST]
This command disconnects the input ports from output ports as specified in
channel_list.
For some modes (such as Single Route Mode, Couple Port Mode, and Bias Mode),
more complex disconnections may occur. For details, refer to Chapter 5.
For example, if Bias Mode is ON, you cannot directly specify the input Bias port.
But if you use :ROUT:OPEN to disconnect a bias enabled output port from another
input port, the output port is automatically connected to the input Bias Port.
Syntax [:ROUTe]:OPEN[:LIST] (@channel_list)
Example OUTPUT @Hp5250;":ROUT:OPEN (@10101,10202)"
Parameter Explanation
card_number card number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
Parameter Explanation
channel_list Channels to open (disconnect).
For details about channel_list, refer to Chapter 5.