Command Reference :DIAGnostic:TEST:FRAMe[:EXECute]?

:DIAGnostic:TEST:FRAMe[:EXECute]?

This command executes the specified E5250A test, then returns the test result. For more info, see Chapter 3.

After :DIAG:TEST:FRAM:EXEC CONT execution, the E5250A status becomes same as after *RST command execution.

After :DIAG:TEST:FRAM:EXEC FPAN execution, the E5250A status is not changed.

NOTE

 

The :DIAG:TEST:FRAM:EXEC FPAN command requires key input from the

 

 

front panel. If key is not pressed within 10 seconds, this test result will be fail.

 

 

:DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller FPANel

Syntax

 

 

 

 

 

 

 

 

 

Parameter

Explanation

 

 

 

 

 

 

CONTroller

Controller test

 

 

 

 

 

 

FPANel

Front panel interface test

 

 

 

 

Query response

 

test_result <newline><^END>

 

1

: Fail

 

 

0

: Pass

 

Example OUTPUT @Hp5250;":DIAG:TEST:FRAM? CONT"