Setting up Measurement Environment

Mounting Connectors Directly

For highly accurate current forcing and measurements while minimizing leakage, surround all force and sense lines from SMU by a guard as far as possible, and make cables stable by taping.

For reducing capacitance measurement error, tape the cables to any grounded materials such as shielding box. If you use probe card, grounding the probe card also reduces capacitance measurement error.

NOTE

Low-Noise Coaxial Cable

When you make a SMU output connection by using low-noise coaxial cable (Agilent part number: 8121-1191), shave the Conductive layer and the Insulator (clear) by knife so that the Center Conductor is insulated from the Conductive layer. Refer to following figure.

 

Cover here using sleeve

 

Insulator (black)

min 2 mm

Center conductor

 

 

 

(for Force/Sense signal)

Outer conductor

 

approx. 10 to 15 mm

Insulator (clear)

(for Guard signal)

Conductive layer (black)

 

 

 

4-14

Agilent E5250A User’s Guide, Edition 9

Page 84
Image 84
Agilent Technologies Agilent E5250A manual Low-Noise Coaxial Cable