VXIplug&play Driver

Driver Functions

hpe5250a_compenC

This function compensates capacitance/conductance data measured by using Agi- lent 4284A C meter, and returns compensation results. If you change the compensa- tion data, create the compensation data file, and specify the data file using hpe5250a_selectCompenFile function before executing this function.

Syntax

ViStatus _VI_FUNC hpe5250a_compenC(ViSession vi,ViReal64 frequency,

 

ViReal64 len_hptrx, ViReal64 len_usrtrx_h, ViReal64 len_usrtrx_l,

 

ViReal64 len_usrcoax_h, ViReal64 len_usrcoax_l, ViReal64 raw_c,

 

ViReal64 raw_g, ViPReal64 compen_c, ViPReal64 compen_g);

Parameters

vi

Instrument handle returned from hpe5250a_init( ).

 

frequency

Measurement frequency. 1E+3 to 1E+6 Hz.

 

len_hptrx

Agilent 16494A triaxial cable. 1.5 or 3.0 m.

 

len_usrtrx_h

Triaxial cable length (in m) between connector plate and DUT

 

 

high terminal. If you do not use triaxial cable, enter 0 (zero).

 

len_usrtrx_l

Triaxial cable length (in m) between connector plate and DUT

 

 

low terminal. If you do not use triaxial cable, enter 0 (zero).

 

len_usrcoax_h

Coaxial cable length (in m) between connector plate and DUT

 

 

high terminal. If you do not use coaxial cable, enter 0 (zero).

len_usrcoax_l

raw_c raw_g compen_c compen_g

Coaxial cable length (in m) between connector plate and DUT low terminal. If you do not use coaxial cable, enter 0 (zero).

Capacitance value (in F) measured by the 4284A.

Conductance value (in S) measured by the 4284A.

Capacitance compensation result (in F).

Conductance compensation result (in S).

8-18

Agilent E5250A User’s Guide, Edition 9

Page 242