Controlling the E5250A

Switch Control Basics

The two output ports are determined by the channel_list parameter of the :ROUT:CLOS command, and the lower output port number is always odd.

So, when you execute :ROUT:CLOS or :ROUT:OPEN command for the couple port, four relays operate synchronously as shown in Figure 5-3.

For example, assume that input ports 1 and 2 are coupled, and that the Couple Port mode is ON:

1.If :ROUT:CLOS command specifies to connect channel 0105, relays are closed to realize two channels (0105 and 0206).

2.If :ROUT:OPEN command specifies to disconnect channel 0105, all four relays between the two input ports and two output ports are opened.

Figure 5-3

Couple Port Relay Connections

2XN-1

INPUT

2XN

2XM-1 2XM

OUTPUT

1. Connect condition

Close

Open

2XN-1

INPUT

2XN

2XM-1 2XM

OUTPUT

2. Disconnect condition

N=1 to 5

M=1 to 6 for Normal Config. M=1 to 24 (max) for Auto Config.

NOTE

For each E5252A installed in the E5250A, you can independently specify the same

 

or different couple ports.

 

 