Executing Sample Programs

Vth and Capacitance Measurement Program

To use the VFP data upload library

If you want to use your VFP data in the program, change the program as follows. Before executing the program, you need to create the VFP data and save it to a file on diskette. And you must link the VFP Data Upload Library to the program. For details, refer to “Using the Virtual Front Panel (VFP) Utility” on page 5-4 and “Using the VFP Data Upload Library” on page 6-22.

1.Insert an exclamation mark (!) at the beginning of each program line shown in the following table, and enter the lines.

Subprogram

Program Line

 

 

Meas_vth

Test_comm$="MOSFET Vth Measurement"

"Constant Definition" block

Channel$="(@10101,10202,10303,10404)"

Meas_vth

OUTPUT @Hp5250;":ROUT:CLOS "&Channel$

"Connect to DUT" block

OUTPUT @Hp5250;"*OPC?"

 

ENTER @Hp5250;A

Meas_vth

OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"

"Connection All Open" block

OUTPUT @Hp5250;"*OPC?"

 

ENTER @Hp5250;A

Meas_c

Test_comm$="Capacitance Measurement"

"Constant Definition" block

Channel$="(@10905,11006)"

Meas_c

OUTPUT @Hp5250;":ROUT:CLOS "&Channel$

"Connect to DUT" block

OUTPUT @Hp5250;"*OPC?"

 

ENTER @Hp5250;A

Meas_c

OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"

"Connection All Open" block

OUTPUT @Hp5250;"*OPC?"

 

ENTER @Hp5250;A

Set5250init

OUTPUT @Hp5250;":ROUT:FUNC NCON"

 

OUTPUT @Hp5250;":ROUT:BIAS:STAT ALL,OFF"

 

OUTPUT @Hp5250;":ROUT:CONN:RULE ALL,SROU"

 

OUTPUT @Hp5250;":ROUT:CONN:SEQ ALL,BBM"

 

OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"

 

OUTPUT @Hp5250;"*OPC?"

 

ENTER @Hp5250;A

Agilent E5250A User’s Guide, Edition 9