Setting up Measurement Environment

Mounting Connectors Directly

To Connect Interlock Circuit to Instrument

Before beginning the measurement, connect the interlock circuit to the interlock connector of the instrument as follows.

For Agilent 4155/4156 semiconductor parameter analyzer:

Connect Intlk connector of the 4155/4156 to the interlock connector by using Agilent 16493J Interlock cable that is furnished with the 4155/4156. You can connect it directly without using any adapter.

For the instruments which has a BNC-Type interlock connector:

If you use an instrument that has a BNC-type interlock connector (such as Agilent 4142B), connect interlock connectors of instrument and the interlock circuit as follows:

Required Parts:

Agilent

Quantity

Description

Model No.

 

 

 

 

 

16435A

1

Interlock Cable Adapter

 

 

 

16493J

1

Interlock Cable

 

 

 

Procedure:

1.Connect interlock connector of instrument to the BNC connector of the 16435A (interlock cable adapter) by using the BNC cable that is furnished with the adapter.

2.Connect the 16435A adapter to the interlock connector by using the 16493J Interlock cable.

Agilent E5250A User’s Guide, Edition 9

4-11