Setting up Measurement Environment

Measurement Cable Length

Table 4-6

Guard Capacitances of E5250A Measurement Environment (Typical)

 

 

 

 

 

 

Connection

Agilent Model/Part No.

Cable

Guard

 

Length

Capacitance

 

 

 

 

 

 

 

 

 

from SMU to

16494A-003 (for non-Kelvin)

80 cm

75 pF

 

E5250A

 

 

 

 

16494B-003 (for Kelvin)

80 cm

90 pF

 

 

 

 

 

 

 

 

 

16494C-001 (for 4142B)

1.5 m

140 pF

 

 

 

 

 

 

E5250A and

E5252A (only 1 card is installed)

145 pF

 

Plug-in Card

 

 

 

 

E5252A (more 1 card)

8 pF/extra card

 

 

 

 

 

 

 

 

 

E5255A (only 1 card is installed)

180 pF

 

 

 

 

 

 

 

E5255A(more 1 card)

60 pF/extra card

 

 

 

 

 

 

from Card

16494A-001 (for non-Kelvin)

1.5 m

125 pF

 

Output to

 

 

 

 

16494A-002 (for non-Kelvin)

3 m

240 pF

 

Connector

 

 

 

 

 

Plate

16494B-001 (for Kelvin)

1.5 m

140 pF

 

 

 

 

 

 

 

 

 

16494B-002 (for Kelvin)

3 m

260 pF

 

 

 

 

 

 

 

16494C-001 (for 4142B)

1.5 m

140 pF

 

 

 

 

 

 

 

16494C-002 (for 4142B)

3 m

260 pF

 

 

 

 

 

 

 

16494D-001

1.5 m

125 pF

 

 

 

 

 

 

 

16494D-002

3 m

240 pF

 

 

 

 

 

 

 

16494E-003

3 m

240 pF

 

 

 

 

 

 

from

Agilent Part No. 8121-1191

X m

130 pF per m

 

Connector

 

 

 

 

Plate to DUT

 

 

 

 

 

 

 

 

4-24

Agilent E5250A User’s Guide, Edition 9

Page 94
Image 94
Agilent Technologies Agilent E5250A manual Guard Capacitances of E5250A Measurement Environment Typical