Executing Sample Programs

Vth and Capacitance Measurement Program

Vth Measurement Setup

Measurement setup of the E5250A and 4155/4156 for this program is as follows:

E5250A/E5252A Setup:

Slot No

Channels

110101, 10202, 10303, and 10404

2not used.

3not used.

4not used.

4155/4156 Setup:

Below is summary of measurement setup saved in E5252.MES file. For details of the setup data, GET the E5252.MES file using the 4155/4156 and refer to the setup displayed on the 4155/4156 screen.

Channel/Source Setup:

 

SMU

V NAME

I NAME

 

MODE

FUNCTION

Source Setup

 

 

 

 

 

 

 

 

 

 

SMU1

Vd

Id

 

V

CONST

5V

 

 

 

 

 

 

 

 

 

 

SMU2

Vg

Ig

 

V

VAR1

0 to 5V, 50mV step

 

 

 

 

 

 

 

 

 

 

SMU3

Vs

Is

 

COMMON

CONST

Common

 

 

 

 

 

 

 

 

 

 

SMU4

Vsb

Isb

 

V

CONST

0V

 

 

 

 

 

 

 

 

 

User Function Setup:

 

 

 

 

 

 

 

 

 

 

 

 

NAME

 

UNIT

 

 

DEFINITION

 

 

 

 

 

 

 

 

SQRTId

 

 

 

SQRT(Id)

 

 

GRAD

 

 

 

DIFF(SQRTId,Vg)

 

 

VTH

 

V

 

@L1X

 

 

 

BETA

 

 

 

@L1G^2

 

 

Agilent E5250A User’s Guide, Edition 9

9-5

Page 261
Image 261
Agilent Technologies Agilent E5250A Vth Measurement Setup, Slot No Channels, Channel/Source Setup, User Function Setup