Executing Self-Test and Leak Test

Using the Self-Test Utility

5.Select the desired SMUm softkey (m=1 to 6). Following message is displayed depending on the connection type:

Message for Kelvin connection:

Disconnect Kelvin cable from IVn terminal.

And disconnect all cables from output terminals.

Message for non-Kelvin connection:

Disconnect cable from IVn terminal.

And disconnect all cables from output terminals.

Where n=1 to 6. And the following softkeys are displayed.

Softkey

Description

 

 

CONTINUE

To continue setting up the Leak Test.

 

 

CANCEL

Cancels the Leak Test execution.

 

 

6.Disconnect all cables from the E5250A input/output terminals. Then connect triaxial cable or Kelvin triaxial cable to the 4155/4156 SMU terminal you specified in step 5 because the next step executes the zero offset cancel.

7.Select the CONTINUE softkey. The following message is displayed depending on the connection type:

Message for Kelvin connection:

Connect SMUm to IVn by Kelvin cable.

Message for non-Kelvin connection:

Connect SMUm to IVn by cable.

Where m=1 to 6 (specified by SMUm softkey) and n=1 to 6 (specified by IVn softkey).

And the following softkeys are displayed.

Softkey

Description

 

 

CONTINUE

Executes the Leak Test. Consecutively connects input port

 

specified by IVn softkey to each output port and makes

 

leak measurement.

 

 

CANCEL

Cancels the Leak Test execution.

 

 

3-18

Agilent E5250A User’s Guide, Edition 9

Page 68
Image 68