Setting up Measurement Environment

Mounting Connectors Directly

To Install LED Circuit

The 4155/4156 semiconductor parameter analyzer's Intlk connector provides the interlock signal and a LED drive signal.

If a LED is connected between pin 4 and pin 5 (or 6) of the interlock connector, the LED lights to indicate high voltage output when more than ±40 V is forced from an SMU in the 4155/4156.

To install LED circuit on your shielding box, do following:

1.Mount LED on your shielding box. For the dimensions of the LED, refer to Figure 4-5.

2.Use wire to connect the LED between pin 4 and pin 5 (or 6) of the interlock connector. Refer to Figure 4-2.

Figure 4-5

Dimensions of LED (Agilent part number 1450-0641)

4-10

Agilent E5250A User’s Guide, Edition 9

Page 80
Image 80
Agilent Technologies Agilent E5250A manual To Install LED Circuit, Dimensions of LED Agilent part number