NOTE

Executing Sample Programs

HCI Measurement Program

HCI Measurement Program

This section explains how to use and modify the HCI measurement sample program. This section consists of the following sections:

“Introduction”

“Adding the Bias Source Control Routine”

“Setting up the Measurement Environment”

“Executing the Program”

“Modifying the Program”

Introduction

The Hot-Carrier-Induced (HCI) degradation of MOSFET parameters is measured by using Agilent 4155/4156 Semiconductor Parameter Analyzer, bias sources, and the E5250A with the E5255A card.

Two HP BASIC language programs are used to perform the HCI measurement:

HCI Test Program (SP_MUX_M) performs HCI degradation test according to the JEDEC proceeding titled "A PROCEDURE FOR MEASURING HCI"

(29-JULY-93).

Analysis Program (SP_MUX_A) determines the lifetime of MOSFET from the data measured by HCI test program.

The SP_MUX_M program flow is shown in Table 9-5.

Table 9-5 also lists the files that are used to set up the 4155/4156 for each part of the SP_MUX_M program. These setup files are stored on the E5250A Program Disk. So, the program disk must be in the 4155/4156's disk drive when you execute the program.

For details about a 4155/4156 setup file, GET the setup file and refer to the setup by using the 4155/4156.

To Add the Bias Source Control Routine

The HCI measurement sample program (SP_MUX_M) requires bias sources. You must modify or add a control routine for your bias sources to the program. Refer to “Adding the Bias Source Control Routine” on page 9-23.

Agilent E5250A User’s Guide, Edition 9

9-17

Page 273
Image 273
Agilent Technologies Agilent E5250A manual HCI Measurement Program, To Add the Bias Source Control Routine