Table 1-4 Standards

This product is designed and manufactured to be in conformity with applicable U.S., Canadian, and International (IEC/CENELEC/CE) standards for intended instrument locations. The following Standards and Specifications are met or exceeded.

Case Protection

IP55 on front door only, when the instrument is panel mounted and the front door

 

securely closed.

 

 

Rear of Panel

EN 60529, IP 20

 

 

Flammability Rating

UL 94 - V2

 

 

Vibration Level

10 to 40 Hz, 0.07 mm displacement; 40 to 60 Hz, 0.2g acceleration

 

 

Electromagnetic

CE EMC Directive 89/336/EEC

Compatibility

 

 

 

Safety

IEC1010 Installation Category II for personal protection

 

 

Intended Instrument

Rack or panel mounting in control room or industrial environments (operator

Locations

accessibility front of panel only)

 

Installation Category II with grounded mains supply from isolation transformer or GFI

 

(ground fault interrupter)

 

Pollution Degree 2 with rear of panel enclosed, in industrial environment

 

 