Online Operation Using Menus

Starting and stopping storage

Data Storage can be started and stopped three ways. At the highest level, the ENABLE STORAGE menu item enables or disables all data storage. At the next highest level, EXTERNAL ENABLE enables or disables storage for each file type (Trends #1-3, Alarms, Events, Diagnostics, Unit Data). At the lowest level, storage for each file type is started and stopped with one of two controls: a menu item BATCH STATE or a discrete parameter BATCH CONTROL.

Table 7-4 describes each method of starting and stopping storage.

 

 

 

Table 7-4 Storage Start/Stop Controls

 

 

Prompt

 

Comments

 

 

 

 

 

 

 

 

 

ENABLE

 

Acts as a master enable/disable switch for all data storage. Select this item from the

 

 

STORAGE

 

DATA STORAGE menu. Set to ENABLE to allow storage of the current schedule. If this

 

 

 

 

is set to DISABLE, online data storage cannot take place, even if storage for

 

 

 

 

individual files is enabled with EXTERNAL ENABLE.

 

 

 

 

 

 

 

EXTERNAL

 

Acts as a enable (high)/disable (low) switch for each storage file type (trends, alarms,

 

 

ENABLE

 

events, diagnostics, unit data). Any storage file whose external enable is configured with

 

 

 

 

a discrete will start and stop according to the state of that discrete. This works

 

 

 

 

independently of batch storage.

 

 

 

 

 

 

 

BATCH

 

Acts as a start/stop switch for each storage file type. The ENABLE STORAGE and

 

 

STATE

 

EXTERNAL ENABLE both must be enabled for BATCH STATE to function.

 

 

 

 

Accessed under DATA STORAGE menu. If batch storage mode is configured for any

 

 

 

 

storage file and if BATCH CONTROL has not been configured with a discrete, then you

 

 

 

 

can manually start/stop the batches using this selection. The instrument will automatically

 

 

 

 

assign a number internally to identify the batch for storage. For example, data stored

 

 

 

 

between the first start and stop is marked as Batch #1. When the next batch starts, the

 

 

 

 

instrument will increment the batch number. These numbers identify the batches for later

 

 

 

 

retrieval.

 

 

 

 

 

 

 

BATCH

 

Accessed under SET UP NEW SCHEDULES. If this item is configured with a discrete, it

 

 

CONTROL

 

will start and stop storage for all storage files configured with batch storage mode. The

 

 

 

 

ENABLE STORAGE and EXTERNAL ENABLE both must be enabled for BATCH

 

 

 

 

CONTROL to function. The instrument will automatically assign a number internally to

 

 

 

 

identify the batch for storage. For example, data stored between the first start and stop is

 

 

 

 

marked as Batch #1. When the next batch starts, the instrument will increment the batch

 

 

 

 

number. These numbers identify the batches for later retrieval.

 

 

 

 

 

 

 

BATCH

 

Accessed under DATA STORAGE. Shows the number (1-255) of the active batch. The

 

 

NUMBER

 

batch number is assigned automatically by the instrument when the batch is started by

 

 

 

 

either method (BATCH CONTROL or BATCH STATE).

 

 

 

 

 

 

 