Language Reference
APBDCTL | Performs trace math DL * (A+B) / C 7 A with each sweep. |
|
| Item |
|
| Description |
| Preset State |
| APBDCTL | OFF | |
Example | 120 | OUTPUT | 723;"SNGLS;" | Activate | |
| 130 | OUTPUT | 723;"LN;" | Select linear amplitude scale . | |
| 140 | OUTPUT | 723;"TS;VIEW TRA;" | Sweep trace A and store it. | |
| 150 | OUTPUT | 723;"APBDCTL;" |
|
Description The APBDCTL command requires a linear display mode. It multiplies the sum of traces A and B by the display line. Then, this result is divided by trace C.
The nal results are stored in trace A. If trace A is in
This command allows ratio measurements to be made as a sweep is taken. If trace B has points near the bottom of screen, overow or poor precision can occur. If a division by zero is attempted, a division by the smallest representable value will be performed instead.