Language Reference

MXM

Item

 

 

 

Description

 

 

Source or Destination Amplitude

0 {

10,000 measurement units

for

linear trace data in

TRA, TRB , TRC,

Range

trace

range , or user-de￿ned trace

. 032,768 { +32,767 measurement

 

units

for logarithmic trace data in

TRA, TRB , TRC, trace range ,

or

 

user-de￿ned trace .

 

 

 

 

 

61.8 2 10308

when using

number , prede￿ned v ariable , or user-de￿ned

 

variable .

 

 

 

 

 

Source-L ength Range

1 {

2048 position

units. Default is

800 position units when using

TRA,

 

TRB, TRC . Truncated if longer than destination length. If shorter than

 

destination length, last element is extended for operation.

 

 

Destination-L ength Range

 

2048 position

units.

 

 

 

 

1 {

 

 

 

 

Example Set the display-line level equal to the larger amplitude of two trace elements.

40

OUTPUT

723;"SNGLS;"

Activate single-sweep mode .

50

OUTPUT

723;"SP

500NM;CENTERWL 1000NM;"

Set measurement range.

60

OUTPUT

723;"TS;"

Sweep measurement

range.

70

OUTPUT

723;"DL

ON;"

Turn on the display

line.

80OUTPUT 723;"MXM DL,AMPU TRA[1],AMPU TRA[400];" Compare the amplitude of the ￿rst and four-hundredth elements of trace A, then set the display line equal to the larger one.

Description The MXM command compares the sources, point by point, sending the greater value of each comparison to the destination. If one of the sources is a single

value, it acts as a threshold and all values greater than the threshold are sent to the destination.

The default is 800 position units when using TRA, TRB, TRC and is truncated if the trace is longer than destination length. If the trace is shorter than the destination length, the last element is replicated to complete the trace.