Language Reference

CALCOR

Individually remove applied calibration correction factors.

 

 

Item

 

 

Description

 

 

 

 

 

 

 

Preset State

 

 

CALCOR

ALL,ON

Example

170

OUTPUT

723;"CALCOR

PWR,?;"

Query status of power correction factor.

 

180

ENTER

723;A

 

Assign response (1 or 0) to computer variable.

Description The CALCOR command disables the correction factors computed by the CAL command. The factors can be disabled or re-enabled individually or all at

once. During normal operation, all correction factors are on. Disable the factors during special applications, or when you wish to supply your own calibration procedure.

CALCOR PWR,OFF disables power correction factors. These factors improve the power accuracy of displayed measurement results by compensating for irregularities in the system.

CALCOR WL,OFF disables wavelength correction factors. These factors improve the wavelength accuracy of displayed measurement results by compensating for irregularities in the monochromator.