Language Reference
CALCOR | Individually remove applied calibration correction factors. |
|
| Item |
|
| Description | |
|
|
|
|
|
| |
| Preset State |
|
| CALCOR | ALL,ON | |
Example | 170 | OUTPUT | 723;"CALCOR | PWR,?;" | Query status of power correction factor. | |
| 180 | ENTER | 723;A |
| Assign response (1 or 0) to computer variable. |
Description The CALCOR command disables the correction factors computed by the CAL command. The factors can be disabled or
once. During normal operation, all correction factors are on. Disable the factors during special applications, or when you wish to supply your own calibration procedure.
CALCOR PWR,OFF disables power correction factors. These factors improve the power accuracy of displayed measurement results by compensating for irregularities in the system.
CALCOR WL,OFF disables wavelength correction factors. These factors improve the wavelength accuracy of displayed measurement results by compensating for irregularities in the monochromator.