Language Reference

SMOOTH

SMOOTH

Reduce amplitude variations on a trace.

Item

 

 

 

 

Description

 

 

 

Number-of-P oints Range

3

to

length

of source

 

 

 

 

 

 

Source-Amplitude Range

0

{

10,000

measurement units

for

linear

trace data.

 

 

 

032,768

{

+32,767 measurement

units

for

logarithmic

trace

data.

Source-L ength Range

3

{

2048

position units. Default

is 800 position

units when

using

TRA,

 

TRB, TRC .

 

 

 

 

 

 

 

Example

5

CLEAR

723

Initialize optical spectrum

analyzer .

7

OUTPUT

723;"IP;"

 

 

10

OUTPUT

723;"VARDEF NN,0"

De￿ne user-de￿ned variable

equal to 0.

20

OUTPUT

723;"CENTERWL 750NM;SP 100NM;"

Set measurement range .

 

30

OUTPUT 723;"REPEAT;";

Begin repeat loop .

 

40

OUTPUT 723;"ADD NN,NN,10;";

Add 10 to NN.

 

50

OUTPUT 723;"SNGLS;CLRW TRA;TS;";

Activate single-sweep mode

. Sweep trace A.

60

OUTPUT 723;"VIEW TRA;";

Store and view trace A.

 

7-441