Language Reference
PDMEAS
Example | 100 | CLEAR 723 | Initialize analyzer . |
| 110 | OUTPUT 723;"IP;" |
|
| 120 | PRINT "STORE REFERENCE TRACE" | Initiate photodiode atness measurement. |
| 130 | OUTPUT 723;"STORREF THRU;" |
|
| 140 | PRINT "HIT RETURN TO CONTINUE" |
|
| 150 | PAUSE |
|
| 160 | OUTPUT 723;"PDMEAS ON;" |
|
Description
See Also
The PDMEAS command is used to measure the atness of an external photodiode. The
When the PDMEAS ON parameter is used, the following conditions are satised:
Trace A 0 B is turned on.
Internal, optical switch is set to external.
Diode atness correction factors are turned o.
External diode is selected.
IN and OUT indicator LEDs are turned on.
DET IN indicator LED is turned o.
When the PDMEAS OFF parameter is USED, the following conditions are satised:
Trace A 0 B is turned o.
Internal, optical switch is set to external.
Diode atness correction factors are turned on.
Internal diode is selected.
IN and OUT indicator LEDs are turned on.
DET IN indicator LED is turned on.
INSTMODE,STORREF