Language Reference

AUTOALIGN

Example

Description

Automatically realign the monochromator output ￿ber.

10 OUTPUT 723;"AUTOALIGN;"

The AUTOALIGN command realigns the optical spectrum analyzer monochromator output ￿ber with the output beam. If there is a marker on screen, the alignment is done at the marker's wavelength. If there is no marker, the alignment is done at the wavelength of the largest signal detected in full span.

The auto-alignment routine is performed in a 10 nm resolution bandwidth setting. At the end of the auto-alignment routine, the instrument returns to its previous state.

See Also ALIGN, ALIGNPRST