Language Reference
TRDSP | Blank display of active trace(s) while measuring. |
Item | Description |
|
|
Preset State | TRDSP TRA,ON; TRDSP TRB,OFF; TRDSP TRC,OFF |
Example | 30 | OUTPUT | 723;"CENTERWL 750NM;SP 100NM;" | Set measurement range. | |
| 40 | OUTPUT | 723;"TS;" |
| Sweep trace A. |
| 50 | OUTPUT | 723;"TRDSP | TRA,0;" | Turn o display of trace A. |
| 60 | OUTPUT | 723;"TRDSP | TRA,ON;" | Display trace A. |
| 70 | OUTPUT | 723;"TRDSP | TRA,?;" | Query |
Description
Query Response
See Also
The TRDSP command turns o trace A, B, or C without clearing the trace data or interrupting measurements. Thus, data continues to be stored in the active trace. The OFF argument reduces the time duration between sweeps.
TRPST, TRSTAT