STORREF

STORREF

Example

Description

Store through references for stimulus-response measurements.

10 OUTPUT 723;"STORREF THRU;"

STORREF THRU discontinues measurements with trace B, then transfers through-reference amplitude information measured with trace A to trace B.

STORREF displays \Not stored; A 0 X 7 A on" and does not operate when the following functions are on:

￿Trace A 0 Trace B (AMB ON)

￿Trace A 0 Trace B + Display Line (AMBPL ON)

￿Trace A 0 Trace C (AMC ON)

If AMB is executed after STORREF, measurements made with trace A are normalized (weighted) according to the contents of trace B. If AMC is executed after STORREF, trace A is normalized according to the contents of trace C.

See Also CLRW, PDMEAS, AMB, AMC, AMBPL

7-469