Language Reference

OPTSW

Example

Description

Query Response

See Also

Activate the front-panel MONOCHROMATOR OUTPUT and PHOTODETECTOR INPUT ports.

10 OUTPUT 723;"OPTSW INT;"

The OPTSW command selects the signal path on HP 71451B instruments (HP 70951B modules). The INT argument selects the normal signal path through the optical spectrum analyzer's monochromator to the photo diode for detection. The EXT argument makes the monochromator's output available at the front panel's MONOCHROMATOR OUTPUT connector. Also, the front panel's PHOTODETECTOR INPUT connector is activated for input.

The correction data is not changed when the switch is changed.

CORSEL, INSTMODE

7-342