Language Reference

WLMKRL

WLMKRL

Position the left power integration point to a location based upon the

 

wavelength value entered.

Example

100

OUTPUT

723;"CENTERWL

900NM;"

Set the center wavelength to 900 nm.

 

105

OUTPUT

723;"TS;"

 

 

Update the current trace .

 

110

OUTPUT

723;"WMLKRR

1000E 09;"

Position the right power integration point

 

 

 

 

 

 

to 1000 nm.

 

12

OUTPUT

723;"WLMKRL

800

NM;"

Position the left power integration point to

 

 

 

 

 

 

800 nm.

Description

See Also

The WLMKRL command positions the left power integration point to a location based upon the wavelength value entered. The left point is speci￿ed as marker number 4 and the right point is speci￿ed as marker number 5.