Language Reference

MKSS

MKSS

Set center-wavelength step size equal to the wavelength di￿erence between

 

delta markers.

Example

30

OUTPUT

723;"STARTWL 600NM;STOPWL 800NM;"

Set measurement

range.

 

 

40

OUTPUT 723;"MKN

EP;"

Move marker to fundamental,

 

 

 

 

 

using

front-panel

knob,

then

 

 

 

 

 

 

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

 

 

 

 

 

 

press

ACCEPT VALUE key.

 

50

OUTPUT

723;"MKCWL;"

Move the marked signal to

the

 

 

 

 

 

center of the display .

 

 

60

OUTPUT

723;"MKD;"

Turn on delta marker.

 

 

65

OUTPUT

723;"MKPK

NH;"

Move the marker to the next

 

 

 

 

 

highest peak.

 

 

 

70

OUTPUT

723;"MKSS;"

Set the center-wavelength

step

 

 

 

 

 

size equal to the delta marker

 

 

 

 

 

wavelength separation.

 

Description

See Also

The MKSS command uses delta marker information to set the center- wavelength step size. Delta markers must be displayed before using this command. MKSS is inoperative when executed after MKAL or MKAR. MKSS sets the center-wavelength step size equal to the wavelength di￿erence of the delta and reference markers.

SS, MKD

7-295