Language Reference
MKPITX
MKPITX | Control the excursion value for marker pit operations. |
| Pit-Excursion Range | | 0.01 { | 300 | dB | |
| Step Increment | | 1 scale | division during log mode. 1 scale division | at top graticule line |
| | | (.915 dB) during linear mode. | |
| Preset State | | MKPITX | 3DB | |
Example | 100 OUTPUT | 726;"MKPITX 6DB;" | Set the pit excursion value | to 6dB. |
Description Just as MKPX sets the peak excursion value for marker peak operations, MKPITX sets the pit excursion value for marker pit operations. The MKPITX
command species the minimum signal excursion for the marker-pit functions (MKPK). Signal movement less than the pit-excursion value is not considered a pit. To be a pit, the skirt of the signal response must dip by at least the pit-excursion value, and then rise by at least the pit-excursion value.
The default pit-excursion value is 3 dB for pits displayed on a logarithmic scale. This means that any pit with an excursion of less than 3 dB on either side of its bandwidth shape is not identied as a pit. If the marker-pit command, MKPK MIPIT , is executed on such a pit, the optical spectrum analyzer does not place a marker at the pit. For signals displayed on a linear scale, the default pit excursion equals
excursion = 1 | 0 10 | MKPITX |
|
200 |
7-285